Raman based pH measurements with polyaniline

    Forskningsoutput: TidskriftsbidragArtikelVetenskapligPeer review

    125 Citeringar (Scopus)

    Sammanfattning

    The reversible acid-base equilibrium between the emeraldine salt (ES) and the emeraldine base (EB) form of thin electropolymerized films of polyaniline (PANI) has been studied between pH 2 and 9 with in situ Raman spectroscopy using the 514, 633 and 780 nm laser excitation wavelengths. The electropolymerization of PANI was done with cyclic voltammetry in a flow cell from a solution consisting of 50 mM aniline and 1.0 M HCl. It is shown that the Raman signal of the C=N stretching vibration at 1439 cm(-1), originating from the quinoid units of the EB form, can be used for pH measurements between pH 3 and 6 with the 633 nm laser. This vibration is strongly resonance enhanced by the 633 nm laser and its intensity grows therefore considerably in the pH interval where the ES-EB transition mainly takes place (pH 3-6). The influence of the film conditioning pH was also studied as well as the reproducibility of the Raman spectra when pH was changed several times from pH 6 to 3.
    OriginalspråkOdefinierat/okänt
    Sidor (från-till)320–329
    Antal sidor10
    TidskriftJournal of Electroanalytical Chemistry
    Volym580
    Nummer2
    DOI
    StatusPublicerad - 2005
    MoE-publikationstypA1 Tidskriftsartikel-refererad

    Nyckelord

    • 514 nm
    • 633 nm
    • 780 nm
    • conducting polymer
    • emeraldine base (EB)
    • emeraldine salt (ES)
    • pH measurements
    • polyaniline
    • Raman spectroscopy

    Citera det här