Sammanfattning
The reversible acid-base equilibrium between the emeraldine salt (ES) and the emeraldine base (EB) form of thin electropolymerized films of polyaniline (PANI) has been studied between pH 2 and 9 with in situ Raman spectroscopy using the 514, 633 and 780 nm laser excitation wavelengths. The electropolymerization of PANI was done with cyclic voltammetry in a flow cell from a solution consisting of 50 mM aniline and 1.0 M HCl. It is shown that the Raman signal of the C=N stretching vibration at 1439 cm(-1), originating from the quinoid units of the EB form, can be used for pH measurements between pH 3 and 6 with the 633 nm laser. This vibration is strongly resonance enhanced by the 633 nm laser and its intensity grows therefore considerably in the pH interval where the ES-EB transition mainly takes place (pH 3-6). The influence of the film conditioning pH was also studied as well as the reproducibility of the Raman spectra when pH was changed several times from pH 6 to 3.
| Originalspråk | Odefinierat/okänt |
|---|---|
| Sidor (från-till) | 320–329 |
| Antal sidor | 10 |
| Tidskrift | Journal of Electroanalytical Chemistry |
| Volym | 580 |
| Nummer | 2 |
| DOI | |
| Status | Publicerad - 2005 |
| MoE-publikationstyp | A1 Tidskriftsartikel-refererad |
Nyckelord
- 514 nm
- 633 nm
- 780 nm
- conducting polymer
- emeraldine base (EB)
- emeraldine salt (ES)
- pH measurements
- polyaniline
- Raman spectroscopy
Citera det här
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver