TY - JOUR
T1 - No difference in implant survival between 28-mm M2a RingLoc metal-on-metal and metal-on-polyethylene total hip arthroplasty
T2 - results from the Finnish Arthroplasty Register
AU - Palomäki, Antton
AU - Hemmilä, Matias
AU - Matilainen, Markus
AU - Eskelinen, Antti
AU - Haapakoski, Jaason
AU - Puhto, Ari Pekka
AU - Kettunen, Jukka
AU - Pamilo, Konsta
AU - Manninen, Mikko
AU - Vasara, Anna
AU - Mäkelä, Keijo T.
N1 - Funding Information:
Ethical approval was granted on June 13, 2017 by National Institute for Health and Wellfare (Dnro THL/926/5.05.00/2017). The authors declare no conflicts of interest. The study was funded by the Finnish Arthroplasty Association and the Finnish Government Research Grant. Data sharing is not possible.
Publisher Copyright:
© 2022 The Author(s).
PY - 2022
Y1 - 2022
N2 - Background and purpose — Long-term outcome of small head (28 mm) metal-on-metal (MoM) total hip arthroplasty (THA) is available mainly for Metasul devices (Sulzer Medica, Winterthur, Switzerland). Biomet MoM THA was frequently used in Finland. Therefore, we assessed long-term survivorship of the M2a 28-mm RingLoc MoM THA (Biomet, Warsaw, IN, USA) and compared it with the metal-on-polyethylene (MoP) RingLoc THA from the same manufacturer. Patients and methods — We conducted a register study based on THAs from the Finnish Arthroplasty Register performed between January 1, 2000 and December 31, 2007. 290 28-mm head M2a MoM THAs and 1,647 28-mm head MoP THAs (reference group) were included. The endpoint was revision for any reason, or revision for aseptic loosening, osteolysis, liner wear, or metallosis as one group. Kaplan– Meier survival estimates were calculated, and revision risks were assessed using a Cox multiple regression model, both with 95% confidence intervals (CI). Results — No difference was found in the 15-year Kaplan–Meier survivorship between the 28-mm head M2a RingLoc MoM THA and the reference group for any reason for revision (87.7% [82.9–92.1] and 83.3% [81.0–85.3], respectively). The adjusted hazard ratio (HR) for any reason for revision for the MoM THA group compared with the reference group was at least equal or better (0.70 [0.48–1.02]). Both groups presented similar survival for revision for aseptic loosening of the cup, osteolysis, liner wear, or metallosis, at 96.2% (92.7–98.0) and 95.4% (93.9–96.5), respectively. Interpretation — In the long-term survival there was no difference between the M2a 28-mm RingLoc MoM THA and 28-mm MoP THA. Further follow-up regimens for M2a 28-mm RingLoc THA patients may be unnecessary, but long-term metal ion and radiological data is needed before any formal suggestions.
AB - Background and purpose — Long-term outcome of small head (28 mm) metal-on-metal (MoM) total hip arthroplasty (THA) is available mainly for Metasul devices (Sulzer Medica, Winterthur, Switzerland). Biomet MoM THA was frequently used in Finland. Therefore, we assessed long-term survivorship of the M2a 28-mm RingLoc MoM THA (Biomet, Warsaw, IN, USA) and compared it with the metal-on-polyethylene (MoP) RingLoc THA from the same manufacturer. Patients and methods — We conducted a register study based on THAs from the Finnish Arthroplasty Register performed between January 1, 2000 and December 31, 2007. 290 28-mm head M2a MoM THAs and 1,647 28-mm head MoP THAs (reference group) were included. The endpoint was revision for any reason, or revision for aseptic loosening, osteolysis, liner wear, or metallosis as one group. Kaplan– Meier survival estimates were calculated, and revision risks were assessed using a Cox multiple regression model, both with 95% confidence intervals (CI). Results — No difference was found in the 15-year Kaplan–Meier survivorship between the 28-mm head M2a RingLoc MoM THA and the reference group for any reason for revision (87.7% [82.9–92.1] and 83.3% [81.0–85.3], respectively). The adjusted hazard ratio (HR) for any reason for revision for the MoM THA group compared with the reference group was at least equal or better (0.70 [0.48–1.02]). Both groups presented similar survival for revision for aseptic loosening of the cup, osteolysis, liner wear, or metallosis, at 96.2% (92.7–98.0) and 95.4% (93.9–96.5), respectively. Interpretation — In the long-term survival there was no difference between the M2a 28-mm RingLoc MoM THA and 28-mm MoP THA. Further follow-up regimens for M2a 28-mm RingLoc THA patients may be unnecessary, but long-term metal ion and radiological data is needed before any formal suggestions.
UR - http://www.scopus.com/inward/record.url?scp=85142884112&partnerID=8YFLogxK
U2 - 10.2340/17453674.2022.5252
DO - 10.2340/17453674.2022.5252
M3 - Article
C2 - 36445034
AN - SCOPUS:85142884112
SN - 1745-3674
VL - 93
SP - 854
EP - 858
JO - Acta Orthopaedica
JF - Acta Orthopaedica
ER -