High-throughput screening of ten microcystins and nodularins, cyanobacterial peptide hepatotoxins, by reversed-phase liquid chromatography-electrospray ionisation mass spectrometry

J. Meriluoto*, K. Karlsson, L. Spoof

*Korresponderande författare för detta arbete

Forskningsoutput: TidskriftsbidragArtikelVetenskapligPeer review

42 Citeringar (Scopus)

Sammanfattning

Liquid chromatography-electrospray ionisation mass spectrometry was evaluated in the high-throughput analysis of microcystins and nodularins, cyanobacterial peptide hepatotoxins. Extracts originating from cyanobacterial strains and field material were separated on a 30 mm x 4 mm I.D. Merck Purospher STAR RP-18e column using a rapid gradient of aqueous formic acid and acetonitrile, ionised by electrospray technique and analysed on a Micromass Quattro II triple-quadrupole mass spectrometer operated in the selected ion recording (SIR) mode. The total analysis time per sample was 2.8 min corresponding to 514 samples a day. The system showed good robustness during a series of 320 repetitive injections of a field sample containing three major microcystins.

OriginalspråkEngelska