Sammanfattning
A procedure for the development of a pen-like, multi-electrode potentiometric sensing platform is described. The platform comprises a seven-in-one electrode incorporating all-solid-state ion-selective and reference electrodes based on the conductive polymer (poly(3,4-ethylenedioxythiophene)(PEDOT)) as an intermediate layer between the contacts and ion-selective membranes. The ion-selective electrodes are based on traditional, ionophore-based membranes, while the reference electrode is based on a polymer membrane doped with the lipophilic salt tetrabutyl ammonium tetrabutyl borate (TBA-TBB). The electrodes, controlled with a multichannel detector system, were used for simultaneous determination of the concentration of Pb2+ and pH in environmental water samples. The results obtained using pH-selective electrodes were compared with data obtained using a conventional pH meter and the average percent difference was 0.3%. Furthermore, the sensing system was successfully used for lead-speciation analysis in environmental water samples. (c) 2010 Elsevier B.V. All rights reserved.
Originalspråk | Odefinierat/okänt |
---|---|
Sidor (från-till) | 199–205 |
Antal sidor | 7 |
Tidskrift | Sensors and Actuators B: Chemical |
Volym | 146 |
DOI | |
Status | Publicerad - 2010 |
MoE-publikationstyp | A1 Tidskriftsartikel-refererad |
Nyckelord
- Water analysis
- Reference electrode
- speciation
- SOLID-CONTACT ELECTRODE