An outer layer of pure silicone rubber (SR), i.e. SR without any plasticizer, ionophore, or lipophilic anion, was applied on top of a conventional poly(vinyl chloride) (PVC) based K+-selective membrane in a solid-contact ion-selective electrode (SC-ISE). The influence of the outer SR coating on the analytical performance of the K+-ISEs was studied. The presence of the SR coating did not affect the selectivity of the SC-ISE, indicating that the plasticizer, ionophore, and lipophilic anion are spontaneously distributed from the PVC-based membrane into the SR layer. This was confirmed by electrochemical impedance spectroscopy (EIS). Interestingly, the reproducibility of the standard potential of the conditioned SC-ISE was significantly improved from E-0 +/- 35.3 mV to E-0 +/- 3.5 mV simply by adding the SR coating on top of the plasticized PVC based K+-selective membrane. Moreover, the adsorption of bovine serum albumin (BSA) was significantly reduced at the SR coated ion-selective membrane. Thus, the addition of a SR coating on a plasticized PVC ion-selective membrane seems to be a feasible method to improve the analytical performance and to reduce the biofouling of potentiometric ion sensors.
| Alkuperäiskieli | Ei tiedossa |
|---|---|
| Sivut | 10524–10531 |
| Sivumäärä | 8 |
| Julkaisu | Analytical Chemistry |
| Vuosikerta | 91 |
| Numero | 16 |
| DOI - pysyväislinkit | |
| Tila | Julkaistu - 2019 |
| OKM-julkaisutyyppi | A1 Julkaistu artikkeli, soviteltu |
Laitteet
-
Åbo Akademi Functional Printing Center
Toivakka, M. (PI), Rosenholm, J. (PI), Anttu, N. (PI), Bobacka, J. (PI), Huynh, T. P. (PI), Peltonen, J. (PI), Wang, X. (PI), Wilen, C.-E. (PI), Xu, C. (PI), Zhang, H. (PI) & Österbacka, R. (PI)
Luonnontieteiden ja tekniikan tiedekuntaLaitteistot/tilat: Laitos
Viittausmuodot
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver