Multiinstrument study of the electrophoretic mobility of quartz

Marek Kosmulski, Edward Ma̧czka, Władysław Janusz, Jarl B. Rosenholm

Tutkimustuotos: LehtiartikkeliArtikkeliTieteellinenvertaisarvioitu

43 Sitaatiot (Scopus)

Abstrakti

The electrophoretic mobility of quartz in 0.01 mol dm-3 NaCl and NaNO3 over the pH range 2-8 was studied using five different commercial instruments. The mobilities over the pH range 4-8 were relatively consistent, but the mobilities over the pH range 2-4 and the position of the isoelectric point IEP varied from one instrument to another. This result suggests that the discrepancies in the apparent IEP of quartz (and other silicas) reported in the literature are due to the instrument artifacts. © 2002 Elsevier Science (USA).
AlkuperäiskieliEnglanti
Sivut99-103
Sivumäärä5
JulkaisuJournal of Colloid and Interface Science
Vuosikerta250
Numero1
DOI - pysyväislinkit
TilaJulkaistu - 2002
OKM-julkaisutyyppiA1 Julkaistu artikkeli, soviteltu

Sormenjälki

Sukella tutkimusaiheisiin 'Multiinstrument study of the electrophoretic mobility of quartz'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

Viittausmuodot