Fast separation of microcystins and nodularins on narrow-bore reversed-phase columns coupled to a conventional HPLC system

Lisa Spoof, Milla Riina Neffling, Jussi Meriluoto*

*Tämän työn vastaava kirjoittaja

Tutkimustuotos: LehtiartikkeliArtikkeliTieteellinenvertaisarvioitu

36 Sitaatiot (Scopus)

Abstrakti

Six modern narrow-bore (50 mm length × 2-2.1 mm internal diameter) reversed-phase HPLC columns were tested in the rapid separation of ten microcystins and three nodularins, cyanobacterial peptide hepatotoxins. The columns had either a sub-3 μm particle size or were of monolithic silica technology for high efficiency and rapid run times. A standard low-pressure gradient HPLC equipment with diode-array UV detector was used for the chromatography. The gradient mobile phase consisted of water and acetonitrile, both acidified with trifluoroacetic acid, and the gradient rise times were 1-4 min. Chromatographic performance was assessed by visual judgement and by calculating parameters such as capacity factors, resolution, peak width, selectivity and peak form. Several chromatographic conditions produced excellent results. We recommend sub-3 μm particle-based or monolithic reversed-phase columns for routine use in microcystin and nodularin HPLC analyses.

AlkuperäiskieliEnglanti
Sivut954-964
Sivumäärä11
JulkaisuToxicon
Vuosikerta55
Numero5
DOI - pysyväislinkit
TilaJulkaistu - toukokuuta 2010
OKM-julkaisutyyppiA1 Julkaistu artikkeli, soviteltu