Calcium-selective electrode based on polyaniline functionalized with bis[4-(1,1,3,3-tetramethylbutyl)phenyl]phosphate

A1 Originalartikel i en vetenskaplig tidskrift (referentgranskad)


Interna författare/redaktörer


Publikationens författare: Lindfors T, Ivaska A
Förläggare: ELSEVIER SCIENCE BV
Publiceringsår: 2001
Tidskrift: Analytica Chimica Acta
Tidskriftsakronym: ANAL CHIM ACTA
Volym: 437
Nummer: 2
Artikelns första sida, sidnummer: 171
Artikelns sista sida, sidnummer: 182
Antal sidor: 12
ISSN: 0003-2670
eISSN: 1873-4324


Abstrakt

A new Ca2+-selecrive polyaoiline (PANI)-based membrane material suitable far all-solid-state-sensor applications is presented. The membrane consists of electrically conducting PANI, bis[4-(1,1,3,3-tetramethylbutyl)phenyl]phosphoric acid (DTMBP-PO4H), dioctyl phenylphosphonate (DOPP) and cationic (tridodecylmethylammonium chloride, TDMACI) or anionic (potassium tetrakis(4-chlorophenyl)borate, KTpClPB) Lipophilic additives. PANI transforms the ionic response to an electronic signal and functions gt the same time as the membrane matrix. PANI is made soluble in dichloro methane (DCM) with the Ca2+-selective DTMBP-PO4H Ligand and membranes,can easily be prepared by drop-casting on glassy carbon (GC) substrates. PAM membranes with three different plasticizer (z) and additive (s) contents were studied (the molar ratios between the plasticizer or additive and DTMBP-PO4H are denoted as z and s, respectively). The best electrode performance, considering sensitivity and selectivity, was obtained for PANI membranes prepared with z = 0.90 and s = 0.10, containing KTpClPB as ionic additive. The Ca2+-sensitivity of this membrane was 27.8 +/- 0.2 mV/log a(Ca)(2+) (10(-1) to 10(-4) M CaCl2, LOD = 8 x 10(-7) M) in 0.1 M NaCl, and the selectivity coefficients (log K-Ca,j(pot)) towards j = Na+, K+, Li+, NH4+ and Mg2+, determined with the fixed interference method,were -3.9 +/- 0.1, -4.0 +/- 0.1, -3.5 +/- 0.1, -3.5 +/- 0.1, and -3.6 +/- 0.1, respectively. On the other hand, practically no redox or pH sensitivity was observed when TDMACI was used as the ionic additive. The PAM membranes were characterized with:potentiometry, electrochemical impedance spectroscopy (EIS) and UV-VIS.


Nyckelord

all-solid-state, calcium, electrically conducting polymer, ion-selective electrode, polyaniline, potentiometry, soluble

Senast uppdaterad 2019-14-12 vid 04:21