Round Robin test on bio-imaging transfer standard for 3D optical profilers

A4 Conference proceedings


Internal Authors/Editors


Publication Details

List of Authors: Nolvi A, Viitala T, Pérez AG, Sandler N, Hæggström E, Bermudez C, Artigas R, Kassamakov I
Editors: Yakov G. Soskind, Craig Olson
Publication year: 2017
Publisher: SPIE
Book title: Proc. SPIE 10110, Photonic Instrumentation Engineering IV
Volume number: 10110
ISBN: 9781510606616


Abstract

A stair case height Bio-Transfer-Standard (BTS), developed and produced at the University of Helsinki (UH), was measured in two laboratories. The Round Robin test aims to determine whether BTS works with different optical profilers in different laboratories. First the artefact was measured at UH using a custom-built Scanning White Light Interferometer. Then BTS was measured at Sensofar-Tech, S.L. using an S-neox-type interferometer working either in Phase Shifting Interferometry mode or in Imaging Confocal Microscopy mode. To remove the influence of system calibration, a method featuring sample shifting and measurement subtraction was used. The BTS features eight lipid bilayer steps that each are 4.6 ± 0.1 nm tall on average. All 30 measurements done by four different operators at the two laboratories agree to within 0.1 nm which agrees with theoretical estimates and with measurements done using a surface plasmon resonance technique. The Round Robin results show the applicability of the newly developed bio-imaging transfer standard for calibrating 3D optical profilers.